Single Phase Multifunction Reference Energy Meter (0.05/0.1) HC3101B
Product Description
HC3101B single-phase reference standard meter adopts the latest international digital electronic technology, with 32-bit ARM high-speed processor as the core, plus various advanced methods such as transformer active compensation technology, 16-bit high-speed AD conversion technology, and digital compensation technology. The instrument achieves a high accuracy better than the relative error of 0.05% RD.
HC3101B single-phase reference standard meter can simultaneously measure and display voltage (V), current (A), active power (W), reactive power (Var), frequency (Hz), power factor (PF), phase angle (θ), Apparent power (VA) and other electrical quantities; perform harmonic analysis on voltage and current input signals, calculate and display harmonic content, initial phase angle, power, etc.; can also output electrical energy pulses (active energy, reactive energy, harmonic energy,etc.) to facilitate inspection or provide standard pulses for other devices. The display adopts liquid crystal display, so the picture is clear, the operation is simple, and the function setting is convenient. It is an intelligent power measuring instrument
HC3101B single-phase reference standard meter has a wide range of signal input and auxiliary power supply, which can be applied to various systems and environments.The meter is equipped with a serial port (RS232), to allow connection and control on PC and all the testing parameters also could be acquired from PC as auto-measurement.
2. Specifications
2.1.1 Accuracy: 0.05/0.1
2.1.2 Input range
2.1.2.1 Voltage
Voltage range: 480V/240V/120V/60V(Manual or automatic switch), the default is 240V
30-480V Input Impedance>600KΩ(Relative n)
2.1.2.2 Current
Single terminal input, current range internal switch: 100A/10A/1A/0.1A
1mA-120A
2.1.2.3 Frequency
45Hz-65Hz
2.1.3 Measuring Accuracy
RD-read/Relative value
Measuring conditions: temperature 20℃±2℃,
Humidity 35%-85%.
2.1.3.1 Voltage measurement
Error E<=±0.05% 30V…480V(Relative measuring value)
Resolution 0.001V
2.1.3.2 Current measurement
Error: E<=± 0.05% 50mA…120A(Relative measuring value)
E<=± 0.1% 1mA…50mA(Relative range ended value)
Resolution 0.0001A
2.1.3.3 Power
2.1.3.3.1 Active power
|Pf|>0.5 E<=± 0.05% 50mA…120A(Relative measuring value)
|Pf|>0.5 E<=± 0.1% 1mA…50mA(Relative range ended value)
2.1.3.3.2 Reactive power
|Pf|>0.5 E<=± 0.05% 50mA…120A(Relative measuring value)
|Pf|>0.5 E<=± 0.1% 1mA…50mA(Relative range ended value)
2.1.3.3.3 Apparent Power
|Pf|>0.5 E<=± 0.05% 50mA…120A(Relative measuring value)
|Pf|>0.5 E<=± 0.1% 1mA…50mA(Relative range ended value)
2.1.3.3.4 Resolution 0.001W/Var
2.1.3.4 Power Factor/Phase angle
2.1.3.4.1 Power Factor
Range(PF) 0-1
Accuracy 0.05%/0.1%RG PF 0.5-1
Resolution 0.0001
2.1.3.4.2 Phase angle(V/A)
Range(θ) 0-360°
Accuracy 0.1°/ 0.2°
Resolution 0.001°
2.1.3.5 Frequency
0.02%/0.05%RD 45-65Hz
Resolution 0.0001Hz
2.1.3.6 Error compute
Constants range for the Tested energy meter:
1…1000000lmp./kwh(kvarh,kVAh)
1…1000000lmp./Ws (Vars,VAs)
Error count formulary:
Error=(Theory value-Actual value)/ Actual value
The theory value is the electric energy constant multiplied by the number of turns and then divided by the constant of the meter under test, and the actual value is the number of pulses actually emitted by the standard meter measured by the standard meter.
Error display range:
-100.00%……99.999%
2.1.3.7 pulse input
TTL input 0…5V
Frequency input max.200Hz
Power supply for the pulse scanning head 5V
2.1.3.8 Frequency output
Basic constants Fh ( 220V,5A ):22kHz
Fl ( 220V,5A ):2.2Hz
Occupy for blank ratio 1 :1
2.1.3.9 Harmonic test
Voltage harmonic content rate HRUh:
HRUh=Uh/U1*100% Among them: Uh is the hth harmonic voltage, U1 is the fundamental wave voltage.
Current harmonic content rate HRIh:
HRIh=Ih/I1*100% Among them: Ih is the hth harmonic current, and I1 is the fundamental current.
The harmonic order is displayed from 0 to 31.
2.1.3.10 Energy output
2.1.3.10.1 Accuracy
Active energy 0.05, 0.1
Reactive energy 0.1, 0.2
Input Range
Rated Voltage 60V/120V/240V/480V
Rated Voltage 0.1A/1A/10A/100A
2.1.3.10.2 Output
TTL Pulse output, load>2KΩ
2.1.3.10.3 Energy constant(10A clamp input)
a. Auto-shift
Power Input(KW/Kvar) | Frequency Output(KHz) |
3.0001-6.000 | 15.0005-30.0000 |
1.5001-3.0000 | 15.0010-30.0000 |
0.7501-1.5000 | 15.0020-30.0000 |
0-0.75000 | 0-30.0000 |
Remark:
①100A input:
Input range selection power is 10 times against corresponding power of the list above.
②1A input:
Input range selection power is 1/10 times against corresponding power of the listed above.
③0.1A input:
Input range selection power is 1/100 times against corresponding power of the listed above.
b. Manual Operation(10A)
Voltage Measure Range | Voltage (Auto -shift) | Current(A) | ||
10 | 5 | 1 | ||
Energy Constant(r/kWh) | ||||
480 V | 320V-600V | 18×106 | 36×106 | 18×107 |
240 V | 160V-320V | 36×106 | 72×106 | 36×107 |
120 V | 80V-160V | 72×106 | 144×106 | 72×107 |
60 V | 20V-80V | 144×106 | 288×106 | 144×107 |
Remark:
①100A input:
Input energy is 1/10 times against corresponding power of the listed above.
②1A input:
Input energy is 10 times against corresponding power of the listed above.
③0.1A input:
Input energy is 100 times against corresponding power of the listed above.
2.1.4 Operating Ambient
Temperature 0℃-45℃
Relative humidity 35%-85%
2.1.5 Temperature Influence(5℃-45℃)
Voltage, Current, Frequency 15ppm/℃
Power/energy 30ppm/℃
2.1.6 Serial Communication Port
RS-232 : Nine cores communication line directly contacted.
Transmission data V,A,W,Var,PF,θ,Hz
Calibration System setting and operation under PC based.
2.1.7 Auxiliary Power
AC 220V+/-15% , 45-65Hz
3. Dimension
External dimensions of HC3101B: Length×Width×Highness 485mm×340mm×135mm